| THMS350V Specifications |
- Temperature Range -196°C* to 350 °C (*with LNP95 Cooling Pump)
- 16 mm X,Y sample manipulation
- Sample area of 22mm diameter
- No PC required (unless using motorized valve)
- Vacuum tested to 10-3mbar
- Can be used with transmitted or reflected light
- Mounts directly to microscope table or substage
- Stage body size - 160 x 80 x 24 mm
- 100 ohm platinum sensor 1/10th din standard class A
- Temperature stability < 0.1°C
- Direct injection of coolant into block
- Highly conductive metal for improved heat transfer
- Maximum heating / cooling rate of 30°C/min
- Response time of <1 second at 5°C/min at 50°C
- Objective lens minimum working distance 4.5mm
- Condenser lens minimum working distance 13.2
| THMS600 Specifications |
- Temperature range -196° to 600°C (cooling option required)
- Up to 150°C/min heating, 100°C/min Cooling (using LNP95 cooling pump)
- Temperature stability <0.1°C
- 16 mm X,Y sample manipulation
- Sample area 22 mm diameter
- Quick-release gas valves for atmospheric control
- 100ohm platinum resistor sensor. 1/10th Din Class A to 0.1°C
- Light aperture - 2.0mm Ø
- Silver heating block for high thermal conductivity
- Direct injection of the coolant into the silver block
- Single ultra thin lid window - 0.17mm
- Objective lens working distance 4.5mm
- Condenser lens minimum working distance 12.5mm
- Range of condenser extension lenses available
- Can be used with all microscope techniques
- Water cooled stage body for high temperature work (>300°C)
- Suitable for Confocal, Laser Raman and X-Ray
- Sample side loading without removing the stage lid